| Products - Sterilisation & Autoclaves - Vacuquick |
 |
Vacuquick
Vacuquick 13-B
Vacuquick 14-B
| Vacuquick 13 - B |
|
| Vacuquick 13-B is a "stand alone" autoclave with integrated water storage tank and patented air cooling system, and is ready for operation without further Installation work. The Vacuquick 13-B is delivered with an external condensate container for the used water. |
|
|
|
| Vacuquick 13-B |
Warm start 0.5 kg of instruments |
Warm start 4.0 kg of instruments |
Warm start 0.7 kg
Textiles |
| |
|
|
|
| Universal-Program wrapped instruments (134°C / 2 bar) |
13 minutes |
19 minutes |
17 minutes |
| Quick-Program„B“ singlewrapped instruments (max. 1,5 kg) (134°C / 2 bar) |
11 minutes |
16 minutes |
15 minutes |
| Quick-Program „S“ unwrapped instruments (134°C / 2 bar) |
7 Minuten* |
10 Minuten* |
- |
| Gentle-Programm wrapped instruments (121°C / 1 bar) |
25 minutes |
33 minutes |
29 minutes |
| Bowie & Dick-test (loaded with test-indicator) (134°C / 2 bar) |
approx. 15 minutes |
- |
- |
| Vacuum-test(without load) |
approx. 16 minutes |
- |
- |

 |
| Vacuquick 14 - B |
|
| Vacuquick 14-B requires a tap-water connection with the advantage that if it is connected to the water treatment system MELAdem 40, it is automatically supplied with high quality demineralized water. If many sterilization cycles are carried out daily, the MELAdem 47 is recommended. The tap-water connection in conjunction with the high-performance vacuum pump has only a very low heat emission. |
|
|
| Vacuquick 13-B and Vacuquick 14-B are "Class B" autoclaves and thereby comply with the highest requirements of the European Standard EN 13060.
Whether sterilizing wrapped instruments or unwrapped material, sterilization is always conducted with the greatest safety. Thanks to its compact exterior dimensions, the Vacuquick autoclaves of the Premium-Class also fit into the openings of older built-in cupboards, so that a "Class B" autoclave can also be installed in these practices in the same familiar spot. |
|
|
| The required documentation of the sterilization cycles is carried out over the built-in network interface. The data can be stored directly on the practice network. It is also possible to print the data with the protocol printer MELAprint 42 or store the data on the MELAflash CF Card. The sterilization records can be read out on every PC with the MELAflash CF Card reader. |
|
|
| Two helix reference block systems, the MELAcontrol and MELAcontrol PRO are available for the verification of successful sterilization in a class B autoclave. |
|
|
| The documentation of the release is carried out with the MELAdoc system, consisting of the documentation pages and the MELAdoc label printer for sterilization packagings. |
|
|
|
| For an optimum loading of the chamber, the Vacuquick is delivered with a tray mount for 4 trays or for 2 half-tray-cassettes. Likewise different sterilization containers are available. 2 pieces of the container 17K fit into the chamber of the Vacuquick. |
|
|
|
| Vacuquick 14-B |
Warm start 0.5 kg of instruments |
Warm start 4.0 kg of instruments |
Warm start 0.7 kg Textiles |
| |
|
|
|
| Universal-Program wrapped instruments (134°C / 2 bar) |
12 minutes |
15 minutes |
14 minutes |
| Quick-Program„B“ singlewrapped instruments (max. 1,5 kg) (134°C / 2 bar) |
10 minutes |
14 minutes |
13 minutes |
| Quick-Program „S unwrapped instruments (134°C / 2 bar) |
7 Minuten* |
9 Minuten* |
- |
| Gentle-Programm wrapped instruments (121°C / 1 bar) |
24 minutes |
28 minutes |
26 minutes |
| Prion-Program wrapped instruments (134°C / 2 bar) |
29 minutes |
33 minutes |
31 minutes |
| Bowie & Dick-test (loaded with test-indicator) (134°C / 2 bar) |
approx. 15 minutes |
- |
- |
| Vacuum-test (without load) |
approx. 16 minutes |
- |
- |

For more information on our Services please Contact Us.
|